Time file event not executed
Applies to
DECADE II
Summary
A timefile has been created in a DECADE II but when starting, some events are executed (but not all).
Cause
Board identifier is not set correctly in time file line. Global parameters (valve control on connector C) are always executed but board specific parameters not (autozero, range, aux, relays …). Make sure that the board identifier (ID) in the event screen is set to the right board ID otherwise the event will not be active. Â
Diagnosis and remedy
Set board ID parameter for each time file line to corresponding board. Board ID:
- 00000Â Â Â no board, board specific parameter not executed
- 00001Â Â Â board 1
- 00010Â Â Â board 2
- 00100Â Â Â board 3
- 00011Â Â Â board 1 and 2
- 00101Â Â Â board 1 and 3Â etc…

Fig. 1. DECADE II display showing the programming page for a timed event. ID is in the top right corner and should be set to the correct board (never 00000).